YOKOGAWA ELECTRIC CORPORATION - DL1540 - Oscilloscope
Manufacturer:
Image 1 of 3
If you have any other photos or manuals for the
YOKOGAWA ELECTRIC CORPORATION DL1540
you can
upload the files here.
Equipment:
DL1540
Date:
1998
Category:
Group:
Sub Group:
Information
Full range 4 channel
Max. 200MS/s (8 bit)
Max. 120k word (DL1540)
Max. 2M word (DL1540L)
Max. 60 displays / sec update rate
Built-in 3.5 inch FDD
Real time printing function
Evaluating magnetic recording devices
Observing modulated signals
Evaluating transient response of a circuit
Measuring a composite TV signal
Trouble shooting electronics or electric devices
■ SIGNAL OBSERVATION USING LONG MEMORY
● Capturing Signals Using Long Memory for Accurate Waveforms
The DL1520/DL1540 can continuously capture signals using a
record length of up to 10k words, and the DL1540L/DL1520L
using a record length of up to 100k words. Within the desired
observation period, the DL1520/DL1540 can thus capture sig-
nals at a time resolution (sampling rate) of 10 times that for sig-
nals captured using a record length of 1k word, and the
DL1540L/DL1520L at a time resolution of 100 times. (You can
use a longer record length when capturing a single-shot signal,
or when using the roll mode.)
●Zoom Function Displays Eight Traces Simultaneously, Zooming
in Quickly and Easily on Waveform Details
Zooming is an important function in a digital oscilloscope with
a long record length. The DL1500 series inherits the popular in-
tuitive, high-speed zooming function of the DL4000 series. The
DL1500 series eliminates problem of which trace to assign the
zoom trace to. This scope allows you to display the captured
trace and the zoomed trace at the same time.
● Recall Any Waveform with The History Memory Function
(DL1540/DL1540L/DL1520L only)
With a conventional digital oscilloscope, the contents of the
memory are updated at every trigger. So once a waveform is up-
dated, there is no way to bring the previous waveform back.
However, the history memory function of the DL1540/
DL1540L/DL1520L stores the last 100 displays.
●Roll Mode for Maximum Sampling Speed of 200 kS/s
Roll mode is ideal for observing low frequency signals of around
10 Hz, which is difficult to do with a conventional analog oscillo-
scope. The roll mode allows a maximum sampling speed of 20 kS/s
for the DL1520/DL1540, or 200 kS/s for the DL1540L/DL1520L
● Screen Update Speed Independent of Number of Traces
Displayed or Waveform Processing Type
As shown in the figure below, it is easy to overlook an irregular
event if the digital oscilloscope has a slow screen update speed.
With most digital oscilloscopes, the screen update speed falls as
the number of waveforms displayed is increased, or when signal
processing such as automatic measurement of the waveform pa-
rameters is taking place.
The DL1500 Series has been designed primarily for measuring
waveforms; hence signals are measured without reducing the
screen update speed even when the number of waveforms dis-
played is increased or signal processing is added.
●A SINGLE OSCILLOSCOPE TO PERFORM EVERYTHING, FROM
PROCESSING TO RECORDING OF THE CAPTURED SIGNAL
(Optional built-in printer is needed to print captured waveforms)
Automatic Waveform Parameter Measurement Function
With this function you can automatically measure the maximum
voltage, frequency and other parameters of a waveform.
2 Manuals
Datasheet
Manual type:
Datasheet
Pages:
4
Size:
147.9 KB
Language:
english
Revision:
Manual-ID:
7001-60E
Date:
January 1998
Quality:
Electronic document, no scan, very well readable.
Upload date:
Dec. 27, 2020
MD5:
f20a35b1-c6fe-2616-1cd9-dc4802d9f1d3
Downloads:
1700
Datasheet
Manual type:
Datasheet
Pages:
5
Size:
705.5 KB
Language:
english
Revision:
Manual-ID:
Date:
January 1998
Quality:
Electronic document, no scan, very well readable.
Upload date:
Dec. 27, 2020
MD5:
954983be-f8c9-2396-b27c-8007cf631da5
Downloads:
937