Keithley - 2425 - Level Meter
Manufacturer:
Equipment:
2425
Date:
1998
Category:
Group:
Sub Group:
Information
Product overview
The SourceMeter combines a precise, low-noise, highly stable
DC power supply with a
low-noise, highly repeatable, high-impedance multimeter. It
has 0.012% basic accuracy
with 51⁄2-digit resolution. At 51⁄2 digits, the SourceMeter
delivers 520 readings/second
over the IEEE-488 bus. At 41⁄2 digits, it can read up to
2000 readings/second into its inter-
nal buffer. The unit has broad source and measurement ranges:
Model 2400:
•
Source voltage from 5μV to 210V; measure voltage from 1μV
to 211V.
•
Source current from 50pA to 1.05A; measure current from
10pA to 1.055A.
•
Measure resistance from 100μΩ (<100μΩ in manual ohms) to 211MΩ.
•
Maximum source power is 22W.
Model 2410:
•
Source voltage from 5μV to 1100V; measure voltage from 1μV
to 1100V.
•
Source current from 50pA to 1.05A; measure current from
10pA to 1.055A.
•
Measure resistance from 100μΩ (<100μΩ in manual ohms) to 211MΩ.
•
Maximum source power is 22W.
Model 2420:
•
Source voltage from 5μV to 63V; measure voltage from 1μV to
63.3V.
•
Source current from 500pA to 3.15A; measure current from
100pA to 3.165A.
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum source power is 66W.
Models
2425 and 2430:
•
Source DC or pulse voltage from 5μV to 105V; measure
voltage from 1μV to
105.5V.
•
Source DC current from 500pA to 3.15A; measure DC current
from 100pA to
3.165A.
•
Source pulse current from 500pA to 10.5A; measure pulse
current from 100pA to
10.55A. (Model
2430 only.)
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum DC source power is 110W.
•
Maximum pulse source power is 1.1kW. (Model
2430 only.)
Model 2440:
•
Source voltage from 5μV to 42V; measure voltage from 1μV to
42V.
•
Source current from 500pA to 5.25A; measure current from
100pA to 5.25A.
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum source power is 66W.
NOTESThe Models 2400, 2410, 2420, 2425, 2430, and
2440 are
Y2K compliant.
Models with a “-C” suffix have the Contact Check option. See
Appendix F.
Some additional capabilities of the SourceMeter include:
Concurrent measurements of all three functions over the
remote interface.
Source-measure sweep capabilities (linear and logarithmic
staircase sweeps, source sweep list of up to 2500 points,
memory sweep of up to 100 instrument setups).
6-wire Ω measurement with programmable I-source or V-source
with V or I clamp.
4-quadrant source and sink operation.
Up to 12 stages of limit testing with a built-in comparator
for pass/fail testing.
Digital I/O for stand-alone binning operations or interface
to component handler.
Programming language and remote interfaces — The SourceMeter
uses the SCPI programming language and two remote interface
ports (IEEE-488/GPIB and RS-232C).
Trigger-Link interface to
Keithley Series 7000 switching
hardware.
Math expressions — Five built-in, up to five user-defined
(bus only).
Reading and setup storage — Up to 2500 readings and seven
setups (five user defaults, factory default, *RST default)
can be stored and recalled.
Closed-cover calibration — The instrument can be calibrated
either from the front panel or remote interface.
User manual
Manual type:
User manual
Pages:
53
Size:
724.0 KB
Language:
english
Revision:
C
Manual-ID:
2400S-903-01
Date:
Quality:
Electronic document, no scan, very well readable.
Upload date:
March 7, 2014
MD5:
3c9ff54e-64ca-0a36-a6a0-83410e1d09e4
Downloads:
2998
User manual
Manual type:
User manual
Pages:
592
Size:
2.5 MB
Language:
english
Revision:
G
Manual-ID:
2400S-900-01
Date:
May 2002
Quality:
Electronic document, no scan, very well readable.
Upload date:
Aug. 23, 2018
MD5:
2a4420ab-c110-6276-b376-316f115a7fa0
Downloads:
1965