SRS Stanford Research Systems - SR530 - Amplifier
Manufacturer:
No picture available!
Maybe you can
upload a pic
for the
SRS Stanford Research Systems SR530 ?
If you have any other manuals for the
SRS Stanford Research Systems SR530
you can
upload the files here.
.
Equipment:
SR530
Date:
1997
Category:
Group:
Sub Group:
Information
General
Power 100, 120, 220, 240 VAC (50/60 Hz); 35 Watts Max
Mechanical 17" x 17" x 5.25" (Rack Mount Included) 16 lbs.
Warranty Two years parts and labor.
Signal Channel
Inputs Voltage: Single-ended or True Differential
Current: 106 Volts/Amp
Impedance Voltage: 100 MΩ + 25 pF, ac coupled
Current: 1 kΩ to virtual ground
Full Scale Sensitivity Voltage: 100 nV (10 nV on expand) to
500 mV
Current: 100 fA to 0.5 μA
Maximum Inputs Voltage: 100 VDC, 10 VAC damage threshold
2 VAC peak-to-peak saturation
Current: 10 mA damage threshold
1 μA ac peak-to-peak saturation
Noise Voltage: 7 nV/√Hz at 1 kHz
Current: 0.13 pA/√ Hz at 1 kHz
Common Mode Range: 1 Volt peak; Rejection: 100 dB dc to 1KHz
Above 1KHz the CMRR degrades by 6 dB/Octave
Gain Accuracy 1% (2 Hz to 100KHz)
Gain Stability 200 ppm/°C
Signal Filters 60 Hz notch, -50 dB (Q=10, adjustable from 45
to 65 Hz)
120 Hz notch, -50 dB (Q=10, adjustable from 100 to 130 Hz))
Tracking bandpass set to within 1% of ref freq (Q=5)
Dynamic Reserve 20 dB LOW (1 μV to 500 mV sensitivity)
40 dB NORM (100 nV to 50 mV sensitivity)
60 dB HIGH (100 nV to 5 mV sensitivity)
Bandpass filter adds 20 dB to dynamic reserve
Line Notch filters increase dynamic reserve to 100 dB
Reference Channel
Frequency 0.5 Hz to 100 kHz
Input Impedance 1 MΩ, ac coupled
Trigger SINE: 100 mV minimum, 1Vrms nominal
PULSE: ±1 Volt, 1 μsec minimum width
Mode Fundamental (f) or 2nd Harmonic (2f)
Acquisition Time 25 Sec at 1 Hz
6 Sec at 10 Hz
2 Sec at 10 kHz
Slew Rate 1 decade per 10 S at 1 kHz
Phase Control 90° shifts
Fine shifts in 0.025° steps
Phase Noise 0.01° rms at 1 kHz, 100 msec, 12 dB TC
Phase Drift 0.1°/°C
Phase Error Less than 1° above 10Hz
Orthogonality 90° ± 1°
Demodulator
Stability 5 ppm/°C on LOW dynamic reserve
50 ppm/°C on NORM dynamic reserve
500 ppm/°C on HIGH dynamic reserve
Time Constants Pre: 1msec to 100 sec (6 dB/Octave)
Post: 1sec, 0.1 sec, none (6 dB/Octave) or none
Offset Up to 1X full scale (10X on expand)
Both channels may be offset
Harmonic Rej -55 dB (bandpass filter in)
Outputs & Interfaces
Channel 1 Outputs X (RcosØ), X Offset, R (magnitude), R
Offset, X Noise, X5 (external D/A)
Channel 2 Outputs Y (RsinØ), Y Offset, Ø (phase shift of
signal), Y Noise, X6 (external D/A)
Output Meters 2% Precision mirrored analog meter
Output LCD's Four digit auto-ranging LCD display shows same
values as the analog meters
Output BNC's ±10 V output corresponds to full scale input,
<1Ω output impedance
X Output X (RcosØ), ±10 V full scale, < 1Ω output impedance
Y Output Y (RsinØ), ±10 V full scale, < 1Ω output impedance
Reference LCD Four digit LCD display for reference phase
shift or frequency
RS232
Interface controls all functions. Baud rates from 300
to 19.2 K
GPIB
Interface controls all functions. ( IEEE-488 Std )
A/D 4 BNC inputs with 13 bit resolution (±10.24 V)
D/A 2 BNC outputs with 13 bit resolution (±10.24 V)
Ratio Ratio output equals 10X Channel 1 output divided by
the Denominator input.
Internal
Oscillator Range: 1 Hz to 100 kHz, 1% accuracy
Stability: 150 ppm/°C
Distortion: 2% THD
Amplitude: 1% accuracy, 500 ppm/°C stability
User manual