The Model 8412B Phase-Magnitude Display is a plug-in display
unit for the HP Models 8407 A and 8410-Series Network
Analyzers. The instrument contains a CRT which displays test
signal amplitude and phase on two separate CRT traces.
1-3. The amplitude and phase display may show signals which
represent various amplitude ratios depending on the
transducer used in the network analyzer. The displayed
quantities may be ratios of voltage, current, impedance, or
an incident-to-reflected signal.
1-4. Amplitude is displayed in decibels/division and phase
is displayed in degrees/division on the
CRT graticule. Various vertical amplifier sensitivity ranges
for both amplitude and phase may be selected. Front panel
phase offset controls allow the phase to be adjusted for the
most convenient display.