Advance Instruments OS250
you can
upload
the files here
.
Information
Information
Add Comment
These terms and conditions contain rules about posting comments. By submitting a comment, you are declaring that you agree with these rules:
- Although the administrator will attempt to moderate comments, it is impossible for every comment to have been moderated at any given time.
- You acknowledge that all comments express the views and opinions of the original author and not those of the administrator.
- You agree not to post any material which is knowingly false, obscene, hateful, threatening, harassing or invasive of a person's privacy.
- The administrator has the right to edit, move or remove any comment for any reason and without notice.
Failure to comply with these rules may result in being banned from further commenting.
These terms and conditions are subject to change at any time and without notice.
Comments
So far im i see that the Diode bridge (MR301) are not rectifying well.. he are producing a AC current on output, and the DC tension is about 30V (instead 155V, even if i adjust R320). But after i replace a Diode Bridge by a new "2W08", after a couples os seconds i have a drop of tension across C315 (155V to 60V).. i need to troubleshoot a little more, but maybe is the C315, TR301 or TR302 that are not healthy.. we will see :) have a good day everyone! and thanks again!