Schaffner CDN 110 Network Analyzer
Manufacturer:
Model:
CDN 110
Date:
unknown
Class:
Type:
SubType:
Coupling Network
Information
The Coupling Network
CDN 110 serves to inject the following
standardized surge pulses from the surge
Generator into the
mains supply to the device under test:
Voltage surge of up to 6 kV which follows the 1.2/50 (js
curve (open-circuit)
Current surge of up to 3 kA which follows the 8/20 ¿/s curve
(short circuit conditions).
The
CDN 110 permits test pulses of up to 6 kV and is proofed
for up to 10 kV.
The unit is used in conjunction with a surge
Generator in
setting up a test rig. Surge pulse tests can be carried out
on equipment, systems and installations that are in
operation. Such tests are prescribed in IEEE 587, ANSI/IEEE
C62.45, ¡EC 801-5, VDE 0109, etc,
The
CDN 110 is designed to be used with an EUT supply of up
to 420 Veff (phase to phase) or 230 Veff (phase to earth) at
16 A. Operation is simpie and designed to be safe.
This coupling unit is built into a table-top housing which
can be adapted for 19s-rack installation by the addition of
the appropriate mounting brackets.
The construction takes into account the relevant
specifications issued by the VDE and lEC concerning personal
safety.
Interlock circuits prevent dangerous voltages from reaching
the connections to the device under test in an unnoticed or
unprotected manner. A prerequisite is the assembly of a
correctly built test rig. In order to make the assembly of
the test rig easier for the user, protective enclosures,
external warning lamps, coupling modules, etc. are available
as
Accessories.
Manual type:
User Manual
Pages:
46
Size:
1.49 Mbytes (1557528 Bytes)
Language:
english, german
Revision:
Manual-ID:
601-113B
Date:
unknown
Quality:
Scanned document, all readable.
Upload date:
2017 05 13
MD5:
b3bee5407e127ba0f6ca6b0b80ed55f8
Downloads:
813
Information
1 First things first: Safety
...
1
2 Introduction
... 3
3 Applications
...
4
4 Configuration
...
5
5 Operation
... 6
5.1 Safety advice
... 6
5.2 Operating elements
... 6
5.3 Wiring of the connectors
...
8
5.3.1 EUTpowerfeed
...
8
5.3.2 Earth connection
... 8
5.3.3 interlock connector
... 8
5.3.4 Interlock connection
... 9
5.3.5 High voltage connections
... 9
5.4 Test rig
... 9
5.4.1 Schematic layout of a test facility
... 10
5.4.2 Coupling modes with the CDN 110
... 11
5.5 Effect on the device under test
... 13
5.6 Installation
... 14
5.7 Operation of the instruments
... 15
6 Maintenance
... 16
7 Technical specifications
... 17
7.1 Electrical specifications
... 17
7.2 Mechanical specifications
... 18
7.3 Operating and display elements
... 18
8 Ordering information
... 19
9 Optional
Accessories
... 19
10 Warranty
... 20
Comments