INTRODUCTION
The
Wavetek Modeis
3003 through
3006 are rugged, completely
solid-state Signal Generators covering the frequency range
of .001 or 1 to 520 MHz. Models
3004 and
3006 cover the
range of .001 to 520 MHz while Modeis
3003 and
3005 cover
the 1 to 520 MHz range. Modeis
3005 and
3006 offer 100 Hz
resolution while on Models
3003 and
3004 the resolution is 1
kHz. The output can be amplitude or frequency modulated and
the level can be set between +13 and -137 dBm.
1.1.1 FREQUENCY CHARACTERISTICS
The frequency of the instrument is set via 6 or 7
front-panel Lever/Indicator switches which yield a
resolution of 1 kHz or 100 Hz. In addition, remote
programmers are available to facilitate semi-automatic
programming of frequency and GPiB programming of both
frequency and output level.
The accuracy of the instrument is based on a
crystal-controlled
Oscillator that serves as a stable
frequency reference that enables the instrument to provide
high-stability signals to an accuracy of 0.001% over its
specified frequency range. This accuracy includes possible
errors due to short-term drift, long-term drift, incidental
FM, and variations due to line voltage changes and
temperature changes. With the FREQUENCY VERNIER out of the
CAL position, the frequency is accurate to 0.001% ±10 k Hz.
The accuracy of the instrument can be Improved by using
either the optional External Reference input, or the
optional High Stability internal Reference. An Auxiliary
RF output option is also available to drive a
Counter. Refer
to Section 1.3 for a listing of options available for each
instrument.
1.1.2 MODULATION
These instruments also feature both internal and external AM
and FM capabilities. The internal and external sources can
operate independently, or may be used simultaneously to
produce complex modulation.
Internal modulation frequencies include 400 Hz, 1000 Hz, and
two preset frequencies between 100 Hz and 10 kHz, selected
by the user.
Modulation is indicated on a front-panel meter calibrated
for full-scale readings of 30% and 100% AM, and 3, 10, 30,
and 100 kHz deviation for FM operation.
1.1.3 OUTPUT LEVEL FEATURES
The output power is indicated on a front-panel meter
calibrated in both dBm and VRMS, A fifteen-position, 10 dB/
step Attenuator used in conjunction with an 11 dB VERNIER
control provides the user with a range of +13 dBm to -137
dBm. A Programmable Attenuator option is available covering
the range of from 0 to 109.9 dB attenuation in .1 dB steps
with +13 dBm as the 0 dB reference level.
The calibrated output of the Instrument is leveled to within
±0.75 dB across the complete frequency range of the instrument.
Reverse power protection is also available as an option.
Manual type:
Service and User Manual
Pages:
236
Size:
12.90 Mbytes (13531543 Bytes)
Language:
english
Revision:
E
Manual-ID:
6510-00-0024
Date:
unknown
Quality:
Scanned document, reading partly badly, partly not readable.
Upload date:
2011 11 06
MD5:
09f1468d084b43d5252089e667a3e1f3
Downloads:
3842
Manual type:
User Manual
Pages:
66
Size:
2.47 Mbytes (2593329 Bytes)
Language:
english
Revision:
Manual-ID:
Date:
unknown
Quality:
Scanned document, reading partly badly, partly not readable.
Upload date:
2018 08 13
MD5:
91a50ae90da3121e3595a6860e47b717
Downloads:
2612
Information
SECTION 1 GENERAL INFORMATION
1.1 INTRODUCTION ... 1-1
1.2 SPECIFICATIONS ... 1-2
1.3 OPTIONS ... ... 1-4
1.4 ACCESSORIES ... . .'
. 1-5
SECTION 2 OPERATION
2.1 INTRODUCTION ... ... 2-1
2.2 MECHANICAL
INSTALLATION ... 2-1
2.3 ELECTRICAL INSTALLATION ... 2-1
2.4 DESCRIPTION OF FRONT
PANEL ... 2-3
2.5 DESCRIPTION OF REAR PANEL ... , . ... 2-3
2.6 INSTALLATION CHECKS ... 2-3
2.7 OPERATING PROCEDURE ... 2-5
SECTION 3 THEORY OF OPERATION
3.1 INTRODUCTION ... 3-1
3.2 OVERALL BLOCK DIAGRAM. ... ... 3-1
3.3 C315-2- METER BOARD ... 3-6
3.4 ' MODULATION BOARD ... 3-9
3.5 DPS2A -
Power supply ... ' ... 3-9
3.6 DELETED ... 3-11
3.7 M9W-SWEEP
Oscillator ... .3-11
3.8 M10W-6/M1QW-8-OUTPUT
Amplifier ... 3-11
3.9 M172-SWEEP DRIVE/DAC ... 3-13
3.10 M29-1 - FM REFERENCE ... 3-13
3.11 M30-1/M30-4 - CRYSTAL REFERENCE, . ... 3-13
o.1 2 M31A - kHz STEPS..
... ... 3-16
3.14 M149- 100 Hz STEPS ... ' ... ■ ...
.3*18.
3.15 M33-1 ■■ NARROW
Oscillator LOCK ... - ... 3-19
3.16 IVI34 - WIDE
Oscillator LOCK.,
... . .3-19
3.17 M115 - DOWN
Converter ... ' ... 3-22
SECTION 4 PERFORMANCE TESTS
4.1 INTRODUCTION ... ' ... 4-1
4.2 FREQUENCY RANGE AND RESOLUTION TEST ... 4-4
4.3 FREQUENCY ACCURACY TEST ... 4-5
4.4 FREQUENCY STABILITY TEST ... 4-6
4.5 OUTPUT LEVEL ACCURACY TESTS ... 4-6
4.6 HARMONICS TEST ... 4-10
4.7 NON-HARMONICS TEST ... 4-11
4.8 RESIDUAL AM TEST ... . .4-12
4.9 RESIDUAL FM TEST ... 4-13
4.10 INTERNAL MODULATION FREQUENCY ACCURACY TEST ... 4-14
4.11 PERCENT AM ACCURACY TEST ... .4-14
4.12 AM BANDWIDTH
TEST ... 4-15
4.13 AM DISTORTION TEST ... 4-16
4.14 FM DEVIATION TEST ... 4-17
4.15 FM BANDWIDTH TEST. ... 4-18
4-16 FM DISTORTION TEST ... 4-19
4.17 IMPEDANCE TEST ... 4-19
4.18 RFI TEST ... 4-20
*** PERFORMANCE TEST RECORD ... 4-23