INTRODUCTION
The Type 79-02A Amplifier is a dual-channel amplifier and is
provided with two basic amplifier sensitivities and
bandwidths for XI and X10 calibrated operation as follows:
100 millivolts/division sensitivity with a dc to 100
megacycle bandwidth.
10 millivolts/division sensitivity with a dc to 90 megacycle
bandwidth.
- Operation of this plug-in unit enables the user to obtain
two vertical signals on the screen simultaneously or to
display either channel individually.
The VARIABLE gain control, located concentrically with the
VOLTS/DIV switch, has a range of greater than 2-1/2 to 1.
This VARIABLE gain control provides continuous overlap
between steps of the attenuator (VOLTS/DIV) switch. A
detented stop position is provided when the VARIABLE gain
control is set to CAL.
A MODE switch is provided which enables the selection of
either Channel 1 or Channel 2 for separate display or the
combination of both channels for dual display. In addition,
ALTernate or CHOPped modes of operation are available. Use
of the Polarity Inversion switch may give Channel 1 minus
Channel 2 or Channel 2 minus Channel 1 presentation when
MODE switch is set to ADDED.
The Type 79-02A Amplifier is intended for use in the Y
cavity of the Type 766H/F Series Oscilloscopes.
2. FEATURES
•100 Me bandwidth at 100 mv/div sensitivity •10 mv/div
sensitivity with 90 Me bandwidth •Solid state circuitry
•Cascading feature converts instrument from dual trace to a
high-gain, single trace unit; 1 mv/div sensitivity with 30
Me bandwidth nominal.
FEATURES (concluded)
• Front-panel GAIN ADJ control is provided for each channel
to normalize gain between channels
• Polarity Inversion switch enables subtraction of two
signals when operating in the Added mode
• Ground position on Input Selector switch provides ground
reference on base line without necessity of grounding input
of instrument
• Provides equivalent dual-beam scope operation when : used
with Fairchild Type 74-17A Automatic Display
Time Base Plug-in unit ;
1 Handbuch
Reparatur und Bedienungsanleitung
Dokumenttyp:
Reparatur und Bedienungsanleitung
Seitenanzahl:
113
Größe:
4,1 MB
Sprache:
Englisch
Revision:
Dokument-ID:
6704 5512
Datum:
Januar 1965
Qualität:
Gescanntes Dokument, teilweise schwer oder teilweise nicht zu lesen.
Upload Datum:
7. Mai 2017
MD5:
a2115b2c-6890-7d97-c598-95e3cfc85130
Downloads:
371