The 895A DC Differential Voltmeter is half-rack size, and is
equipped with resilient feet and tilt-up bail for field or
bench use. A single instrument may be mounted in a standard
19 inch rack with metal handle-rack adapter kit 881A-102.
Two half-rack size instruments may be mounted side-by-side
with metal handle -rack adapter kit 881A-1Q3.
The 895A can be used as: (a) A conventional voltmeter for
rapid determination of dc voltages from 0 to ±1100 volts to
within ±3% of range setting, (b) A differential voltmeter
for precise measurement of dc voltages of 0 to ±1100 volts
to within ±(0. 0025% of input +0.0001% of range + 5 uv). (c)
A ratiometer for accurately determining the ratio of two
voltages, and (d) A megohmmeter for measurement of
resistance from 0.1 megohm to 10,000,000 megohms with a
typical accuracy of ±5%. The instrument can be used to
indicate voltage excursions about a nominal value. One
feature that should be emphasized is that no current is
drawn from the unknown voltage source, when the voltmeter is
at null, at up to 1100 volts. Thus, the determination of the
unknown potential is independent of its source resistance.
1-4. Leakage resistance of the 895A to ground is 100
megohms. The instrument contains a polarity switch for
convenience in measuring positive or negative voltages, or
in using the instrument as a ratiometer. A grounded recorder
output is included for monitoring the voltage being
measured. Shock, vibration, humidity, and tem -
perature testing verify a design that will provide years of
hard use under severe environmental conditions.
When used as a differential voltmeter, the 895A operates on
the potentiometric principle. An unknown voltage is measured
by comparing it to a known reference voltage with the aid of
a null detector. An accurate standard for measurement is
provided by an 1100 volt reference supply containing a pair
of temperature-controlled zener diodes. A five-dial
Kelvin-Varley voltage divider attenuates the reference
voltage as necessary to obtain a null condition. The unknown
voltage is then indicated by the setting of the voltage
readout dials.
1 Handbuch
Reparatur und Bedienungsanleitung
Dokumenttyp:
Reparatur und Bedienungsanleitung
Seitenanzahl:
73
Größe:
5,0 MB
Sprache:
Englisch
Revision:
Dokument-ID:
Datum:
Juni 1981
Qualität:
Gescanntes Dokument, alles ist lesbar.
Upload Datum:
16. Januar 2018
MD5:
b9086d34-fefe-6bab-7c28-c8321b9d8365
Downloads:
403