General Radio Company - 1863 - Analog multimeter
Manufacturer:
Image 1 of 1
If you have any other photos or manuals for the
General Radio Company 1863
you can
upload the files here.
Equipment:
1863
Date:
1973
Category:
Group:
Sub Group:
Information
The Type 1863 Megohmmeter indicates directly on the panel
meter any resistance from 0.5 to 20,000,000 MO; the Type
1864 (Figure 1-1) indicates resistance from 0.5 to
200,000,000 MO. These ranges are suitable for leakage-
resistance measurements of most types of insulation used in
electrical machinery, electronic devices and components, etc
(Section 4). The voltage applied to the unknown can be 50,
100, 200, 250 or 500 V from the 1863, as selected by the
TEST VOLTAGE switch on the front panel. The 1864 has a
voltage range from 10 to 1090 V that can be set in
1-V steps from 10 to 109 V, and 10-V steps from 100 to 1090
V by the TEST VOLTAGE switch on the front panel.
The 100-volt level is the EIA standard for measurement of
composition, film, and wire-wound resistors above 100
kiloohms. The 500-volt level is a standard value in the
measurement of the insulation resistance of rotating
machinery, transformers, cables, capacitors, appliances, and
other electrical equipment.
Regulated power supply and charging circuit permit rapid and
accurate measurement of the leakage resistance of capacitors.
Guard and ground terminals permit measurement of grounded or
ungrounded two or three-terminal resistors.
A panel warning light indicates when voltage is applied to
the test terminals and thus permits connections to be made
safely
1 Manual
Service and user manual
Manual type:
Service and user manual
Pages:
44
Size:
17.5 MB
Language:
english
Revision:
Manual-ID:
1863-0100-B
Date:
March 1973
Quality:
Scanned document, all readable.
Upload date:
Dec. 30, 2016
MD5:
be7a4168-de09-820f-c908-a7ceb8b3ed36
Downloads:
970